Cambridge Sensors microdot Spécifications Page 13

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 64
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 12
9
Important Information about the Control
Solution
®
Use only microdot Control Solution.
Check the expiration date on the control
solution vial. Do not use if expired.
Control solution, meter, and test strips
should come to room temperature 20° to
25°C (68° to 77°F) before testing.
Shake the vial, discard the first drop of
®
microdot Control Solution, and wipe off the
dispenser tip to ensure a good sample and
an accurate result.
Use only for three months after first opening.
Record the discard date (opening date plus
three months) on the control solution vial.
Discard after three months.
Store the control solution tightly closed at
temperatures below 30°C (86°F). Do not
refrigerate.
Caution: The control solution range printed
®
on the test strip vial is for microdot Control
Solution only. It is used to test meter and test
strip performance.
It is not a recommended range for your
blood glucose level.
Vue de la page 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 63 64

Commentaires sur ces manuels

Pas de commentaire